Interface – Heidolph LABOROTA 20 compact and safety User Manual

Page 101

Advertising
background image

101

E

INTERFACE

28. Connect interface

Make sure the Unit is OFF before you attempt to connect the interface cable.
Use a cable with SUB-D 9 connectors and connect with the terminal connector located
underneath the LABOROTA 20 control control panel.

29. Interface parameters

9600 Baud, no parity, 8 bits, 1 stop bit, no protocol

30. RS 232 interface

This interface allows you to specify all Set-values, move the lift up and down, open and close

aerating valve and vacuum valve.

All Act-values and status of safety functions may be recalled.


31. RS 232 interface commands

query

Command

action

remarks

Act commands

VAC?

Act Vacuum

VAP?

Act Vapor

TAU?

Act T auto

BATH?

Act Bath

CHILL?

Act Chiller

SPEED?

Act Speed

TIME?

Act Time

elapsed time, continuous operation

COUNT?

Act Time

time, timer mode

SET1?

query Set-value

SET2?

query Act-value

Safety functions

SC?

Safety Cover

0 = action; 1 = no action

FS?

Flask Support

0 = action; 1 = no action

RS?

Rotation Stop

0 = action; 1 = no action

OP?

Over

Temperature
Prot.

0 = action; 1 = no action