5 blockage test – INFICON Sensistor ILS500F Leak Detection System User Manual
Page 45
Advertising

45
Recipes
ninp69e1-a (1410)
7.3.5
Blockage Test
External Fill Regulation
If selected, this is the setpoint of fill pressure alarm.
Internal pressure regulation is disengaged and
pressure will be that of the gas supply line. ILS500
F checks that fill pressure is above Pressure
Setpoint before proceeding to gas test step.
Pressure Unit
Select desired unit.
Notice
This test can only be performed if both test ports are used and connected
on either side of the possible blockage.
Blockage Test
Blockage Test Pressure
Blockage Test Time
0.50
2.0
bar
s
Blockage Test Pressure
Minimum pressure to be attained at Test Port 2
during Blockage Test time.
Blockage Test Time
Time within which Blockage Test Pressure must be
attained at Test Port 2.