Troubleshooting

8.1

Fault Symptoms

8.2

Perform Hardware Test

For troubleshooting and testing of the system, use Service menu.

For remote troubleshooting, use Service Run menu.

Venturi Pump and all Gas Valves can be tested automatically.

Fault Symptom

Fault

Measures

Evacuation Failed

Failed to reach vacuum
within the specified time.

Large leak on Test Object
or connections.

Check the compressed air
supply.

Gas Fill Failed

Failed to fill to the right
pressure within the
specified time.

Large leak on Test Object
or connections.

Check the incoming gas
pressure.

Gas Refill Failed

Failed to refill the object.

Large leak on Test Object
or connections.

Check the incoming gas
pressure.

Gas Evac Failed

Failed to reach vacuum
within the specified time.

HW Error During Test

Serious error has occurred
during test.

Check external equipment,
e.g Active Probe.

Test Timeout

Maximum time for the test
was exceeded.

Check that time is correctly
set.

Notice

Before performing the hardware test, carefully check that your tracer gas
and compressed air feed pressures are correct. Wrongly set pressure can
cause erroneous test results.

Hardware Test

OUTPUTS

ILS500 F plugged

ILS500 F + Object

TEST

STOP

Press “TEST”

0.00

0.00

Vacuum:

Pressure: