5 replacing gas valves – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 67

Advertising
background image

67

Maintenance Instructions

ninp69e1-a (1410)

9.2.5

Replacing Gas Valves

1

Use a 3 mm Allen key to remove the four screws holding the valve to be changed.

2

Lift the old valve out and put the new valve in. Notice the correct orientation in the
picture below.

3

Tighten the screws 2-3 mm (0.08-0.12 in.) at a time moving the key from screw to
screw so that the valve doesn’t tilt much.

4

Tighten the screws and replace the cover.

5

Run through the hardware test again to test that the changed/removed valve(s)
perform as required.

6

Use hand probe to check that there is no external leakage (this part of Hardware
Test is routine).

Advertising
This manual is related to the following products: