1 hardware error messages – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 59

Advertising
background image

59

Troubleshooting

ninp69e1-a (1410)

8.2.1

Hardware Error Messages

Error Message

Reason for Error

Corrective Action*

Hardware Error

Vacuum Sensor Error

No power to vacuum
sensor.

Check cable to sensor.

Sensor not connected to
AD.

Check connection to AD.

Damaged vacuum sensor. Send in for repair.

Hardware Error

Pressure Sensor Error

No power to pressure
sensor.

Check cable to sensor.

Sensor not connected to
AD.

Check connection to AD.

Damaged pressure sensor. Replace sensor.

Analog Inputs Power Off

No power to AD module.

Check power cable on left
side of AD module.

*

Contact your supplier if the suggested action does not clear the error.