Troubleshooting

56

ninp69e1-a (1410)

Manifold Tightness (gross)

The overall tightness of the manifold is tested using the pressure decay method. This is
a complement to the gas test, revealing leakage out, through the exhaust etc.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

False pressure and vacuum decay rejects

Increased tracer gas consumption

Evacuation Valve

This step tests that evacuation valve opens to release tracer gas to exhaust. Same test
as previously but under pressure instead of vacuum.

Possible results:

Valve works

Faulty!

Malfunction will result in:

Failure to terminate test cycle

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Manifold Tightness

0.46

Internal Leakage

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Evacuation Valve

0.00

Valve works!