Maintenance Instructions

62

ninp69e1-a (1410)

9

Maintenance Instructions

There are three different parts that needs regular maintenance:

Venturi Pump
Needs regular cleaning.

Gas Valves
Needs regular cleaning and wears out.

Pilot Valves
Maintenance free if incoming compressed air is dry and filtered to 5

μ

m.

Changing Venturi Pump and all Gas Valves takes less than 15 minutes.

9.1

Maintenance Plan

*Depends on the amount of particulates in the objects tested. Metal burrs and other
sharp particles will wear the valves down, requiring shorter maintenance intervals.

Part

Interval

Action

Venturi Pump

3 months

Perform a Hardware Test.

Check Ultimate Vacuum.

Clean venturi nozzles when necessary.

Evacuation, Fill and
Test Port 2 Valves

3-6 months*

Perform a Hardware Test.

Check condition of valves.

Replace or clean valves when necessary.

Vacuum Sensor Valve 12 months

Perform a Hardware Test.

Check condition of valve.

Replace or clean valve when necessary.

Pilot Valves

12 months

Change valve if unexpected pressure
builds.