5 blockage test – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 45

Advertising
background image

45

Recipes

ninp69e1-a (1410)

7.3.5

Blockage Test

External Fill Regulation

If selected, this is the setpoint of fill pressure alarm.
Internal pressure regulation is disengaged and
pressure will be that of the gas supply line. ILS500
F checks that fill pressure is above Pressure
Setpoint before proceeding to gas test step.

Pressure Unit

Select desired unit.

Notice

This test can only be performed if both test ports are used and connected
on either side of the possible blockage.

Blockage Test

Blockage Test Pressure

Blockage Test Time

0.50

2.0

bar

s

Blockage Test Pressure

Minimum pressure to be attained at Test Port 2
during Blockage Test time.

Blockage Test Time

Time within which Blockage Test Pressure must be
attained at Test Port 2.

Advertising
This manual is related to the following products: