Setup

ninp69e1-a (1410)

4.3

Pneumatic Connections

4.3.1

Connecting Compressed Air

1

Use the hose to connect the compressor and the ILS500 F.

4.3.2

Connecting Tracer Gas

Caution

Make sure that compressed air is dry, well filtered and oil free. Recommended filter
grade is 5

μ

m or finer. Inadequate filtering will result in increased maintenance.

Caution

Make sure to use adequate pressure and flow. For more information, see on page
72
.

Warning

Pressurizing objects at too high pressures can result in a burst object. This in turn
can result in serious injury or even death. Never pressurize objects that have not
previously been burst tested or otherwise approved for the chosen test pressure.

Warning

Tracer Gases can be flammable or asphyxiating. Use only ready-made Tracer Gas
mixtures.

Advertising