Pressure decay test, 4 tracer gas filling – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 44

Advertising
background image

Recipes

44

ninp69e1-a (1410)

Pressure Decay Test

7.3.4

Tracer Gas Filling

Pressure Stabilisation Time

Pressure Decay Test Time

Pressure Decay Limit

5.0

5.0

0.10

s

s

bar

Gas Pressure Decay Test

Pressure Stabilisation Time

Delay time before Pressure Decay test begins.

Pressure Decay Test Time

Time during which pressure drop is recorded.

Pressure Decay Limit

Allowed pressure drop during test time.

Warning

The ILS500 F must never be introduced to pressures higher than that approved for
the object to be tested and never beyond the ILS500 F specification.

Notice

Ensure that the test object has time to become filled before Fill Time Out
expires. In particular long narrow objects, as pipes, may need long filling
time.

Tracer Gas Filling

Fill Setpoint

Fill Timeout

External Fill Regulation

Pressure Unit

5.00

10.0

bar

bar

s

Fill Setpoint

Desired tracer gas fill pressure.

Fill Timeout

Object will be rejected if Pressure Setpoint has not
been attained within this time.

Cancels the fill if the test object has a major leak,
opens, or if there are loose connections.