10 leak detector – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 81

Advertising
background image

Technical Data

81

ninp69e1-a (1410)

11.4.10 Leak Detector

Purpose:

Emergency stop interface.

Pin

Signal

Type

Load

Comment

1

+24 VDC

SUPPLY

2.5 A

2

AUX1

-

+/-1-5 A*

Terminal 1 of safe relay contacts for
auxiliary external use.

3

AUX2

-

+/-1-5 A*

Terminal 2 of safe relay contacts for
auxiliary external use.

4

ESTATUS

OUT

0.5 A

Internal emergency circuit stopped.
Use for reset lamp or PLC
monitoring.

5

SAFESPLY**

SUPPLY

-2.5 A

24 VDC supply from EXTERNAL
emergency stop circuitry.

6

COM

GND

1.0 A

Common GND.

*

250 VAC 5 A cosj =1

30 VDC 5 A L/R = 0 ms

240 VAC 2A cosj = 0.3

24 VDC 1A L/R = 48 ms

** SAFESPLY feeds risk associated loads inside the ILS500 F. These include all gas

and tooling valves.

Connector:

9 pin male D-sub.

Purpose:

Connection of external leak detector (ISH2000 or T-Guard)

Cable ISH2000: Pin-to-pin cable and converter.

Cable T-Guard:

Nullmodem cable and converter.

Baudrate:

115200(ISH2000)/19200(T-Guard)

Pin

Signal

1

Not used

2

TD

3

RD