4 status output, 5 tooling interface – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 78

Advertising
background image

78

Technical Data

ninp69e1-a (1410)

11.4.4

Status Output

11.4.5

Tooling Interface

Connector:

6 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.

Purpose:

Test Status Outputs. Sourcing 24 VDC transistor outputs.

Pin

Signal

Type

Load

Comment

1

RUNNING

OUT

0.5 A

Cycle running.

2

ACCEPT

OUT

0.5 A

Tested part accepted.

3

REJECT

OUT

0.5 A

Tested part rejected.

4

ERROR

OUT

0.5 A

Summing error.

5

EOT/FILLED

OUT

0.5 A

End of test or gas filled indicator
(selectable).

6

COM

GND

-2.0 A

Common GND.

Notice

Gas filling status is available on the STATUS connector (pin 5) on the back
of the unit. Connect to a lamp for easy notification of “End of test” status.

Connector:

8 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.

Purpose:

Electrical tooling interface.

Advertising