Technical Data

ninp69e1-a (1410)

11.4

Interfaces and Connectors

All interfaces signals except the serial. Communication interfaces are discrete 24 VDC
logic signals.

Output signals (OUT) are sourcing transistor outputs. Input signals (IN) are transistor
inputs.

Max current of each signal is given in the tables below. Total current (sum) must,
however, be within instrument specification.

11.4.1

Printer Port/RS232

Caution

Outputs are not relay types. Do not connect external drive source such as 24 V or 100/
230 VAC.

Connector:

9 pin male D-sub

Purpose:

Connection of serial printer or logging device (e.g. PC or PLC)

Cable:

Standard female to female file transfer cable (null modem)

Baud Rate:

9600 default (1200 - 115200 selectable)

Pin

Signal

Specification

1

Not used

Standard

RS232C

2

RD

Data rate

9600 baud

3

TD

Data bits

8

4

Not used

Stop bits

1

5

SG

Parity

none

6

Not used

Flow ctrl

none

7

Not used

8

Not used

9

Not used