Technical Data

ninp69e1-a (1410)

Commands

The printer port can also be used to control the ILS500 F. The most commonly used
functions can be started/configured over the RS232 interface. Always use New Line
(0A,LF) as end character.

11.4.2

Input 1 (Optional)

Command

Action

K<0A>

Starts a calibration. If the function is busy, the ILS500 prints the
time remaining for the ongoing calibration in seconds. For
example, if two seconds remains the printout is as follows:

WAIT 2 <0A>

S<0A>

Statistics (see table below)

RS<0A>

Reset statistics

R<09>

Factory Default <0A> loads a recipe. For example
“R<09>Factory Default” loads the recipe Factory Default. When
the recipe is loaded the recipe name is echoed back. If a recipe
name isn’t in the ILS500, the answer from the ILS500 will be
“Not a recipe name!”

Statistics

Printed data

Explanation

REC:AP29

- recipe name.

Printed if recipes is activated

TOT:00031

- total

ACC:00009

- accepted

REJ:00022

- rejected

EVA:00001

- evacuation

VDE:00000

- vacuum decay

BLO:00006

- blockage test

FIL:00001

- gas filling

PRE:00000

- pressure decay

GAS:00014

- gas detector