3 connecting exhaust to air vent, Exhaust recommendation – INFICON Sensistor ILS500F Leak Detection System User Manual

Page 23

Advertising
background image

23

Setup

ninp69e1-a (1410)

4.3.3

Connecting Exhaust to Air Vent

Exhaust Recommendation

The exhaust gas must be directed out of the building.
It is best placed on the roof of the building, far away from the fresh air intake of the
test station.

It is recommend that a dedicated duct is installed. Install an electric duct fan and an
optional wind extractor.

It is not recommended to use the general ventilation system to ventilate the exhaust.
If the ventilation system is equipped with energy recirculation there is a big risk that
large amounts of tracer gas will be carried back to the test room thus disturbing the
testing.

1

ILS500 F

2

Exhaust Hose

3

Bleed Air

2

3

1

Max 10 m (30 ft.)

Ø

25 mm (1 in.)

Ø

100 mm (4 in.)

Min 2 m (6.5 ft.)

Notice

Inadequate exhaust installation is the most common reason for problems
with tracer gas leak testing.

Too narrow or too long exhaust line will result in reduced evacuation
capacity and thereby increased cycle time.