Evacuation valve, Venturi pump – INFICON Sensistor ILS500F Leak Detection System User Manual
Page 52
Advertising

Troubleshooting
52
ninp69e1-a (1410)
Evacuation Valve
Evacuation valve is checked for internal leakage.
Possible results:
•
No Internal Leakage
•
Internal Leakage
Internal leakage can result in:
•
False vacuum decay rejects
•
Increased tracer gas consumption
Venturi Pump
Checking max vacuum of Venturi pump.
Possible results:
•
Max Vacuum OK
•
Poor Max Vacuum
Poor max vacuum can result in:
•
Failed pre-evacuation
•
Slower evacuation
Hardware Test
CONTINUE
STOP
Vacuum:
Pressure:
-0.03
0.00
bar
bar
Evacuation Valve
No Internal Leakage
Hardware Test
CONTINUE
STOP
Vacuum:
Pressure:
-0.88
0.00
bar
bar
Venturi Pump
Max Vacuum OK