1 hardware error messages – INFICON Sensistor ILS500F Leak Detection System User Manual
Page 59
Advertising

59
Troubleshooting
ninp69e1-a (1410)
8.2.1
Hardware Error Messages
Error Message
Reason for Error
Corrective Action*
Hardware Error
Vacuum Sensor Error
No power to vacuum
sensor.
Check cable to sensor.
Sensor not connected to
AD.
Check connection to AD.
Damaged vacuum sensor. Send in for repair.
Hardware Error
Pressure Sensor Error
No power to pressure
sensor.
Check cable to sensor.
Sensor not connected to
AD.
Check connection to AD.
Damaged pressure sensor. Replace sensor.
Analog Inputs Power Off
No power to AD module.
Check power cable on left
side of AD module.
*
Contact your supplier if the suggested action does not clear the error.