6 replacing pilot valves – INFICON Sensistor ILS500F Leak Detection System User Manual
Page 68
Advertising

Maintenance Instructions
68
ninp69e1-a (1410)
9.2.6
Replacing Pilot Valves
1
Use small screw driver to loosen the screw holding the valve. You must back the
screw all the way out until you feel it “jumping” in the thread entrance.
2
Push down on the LEDs while pressing the screw down until you feel the locking
mechanism “snap”.
3
Lift the old valve out from the coil side. If the valve does not come off, repeat steps
3 and 4 making sure the screw is completely backed out.
4
Push the screw in while inserting the new valve. Insert the end facing the screw
first and then push the coil side down.
5
Tighten the screw.
6
Replace the cover.