Chapter 3 SQS-242 Software

3-10

Rate Ramps Tab

Rate ramps cause changes to the deposition rate over time under PID control. Each
rate ramp has a starting thickness, an elapsed time to ramp to the new rate, and a new
rate setpoint. Each process layer can have an unlimited number of rate ramps.


Insert Ramp: Inserts a new rate ramp for the selected layer, at the selected position in
the rate ramps list. Existing rate ramps are shifted down.

Delete Ramp: Deletes the selected rate ramp.

Move Up: Shifts the selected rate ramp up one position.

Move Down: Shifts the selected rate ramp down one position.

Start Thickness: The thickness that triggers a timed ramp to a new rate. (Start
thickness should be greater for each subsequent ramp, and less than the final layer
thickness, otherwise the rate ramp is skipped.)

Ramp Time: The time (in seconds) to ramp to the new rate. If the rate ramp is too fast,
a PID control error may be generated.

New Rate: The new deposition rate setpoint for the selected layer.