Chapter 3 SQS-242 Software

3-19

Sources Tab


Output Name
: A meaningful name assigned to SQM-242 outputs 1 and 2 in single card
systems; 1 to 4 in two card systems. For clear display, keep the name to less than 8
characters.

Full Scale: The input voltage required by the deposition source power supply to
produce 100% output power. Positive or negative full scale values are possible.

Completion Timer: If Complete Signal is checked, the process will halt if a move
complete signal is not received within this timeout period. If Complete Signal is not
checked, the process waits for this fixed time period before starting a layer.

Complete Signal
: Check this box if your source index sends a signal designating that
the move is complete and preconditioning can begin.

Test Output: Useful for testing source wiring and Full Scale voltage settings. Select an
output, then click Send to set the SQM-242 card output to its Full Scale voltage. Click
Clear to return the selected output to 0 volts.