Chapter 3 sqs-242 software – INFICON SID-242 Thin Film Deposition Controller User Manual

Page 44

Advertising
background image

Chapter 3 SQS-242 Software

3-14

Source/Sensor Tab

The Source/Sensor tab controls the physical setup of the deposition system.


Material: Selects the physical deposition material

Max Power: The maximum output power allowed for the selected output. The full scale
output voltage is a function of the deposition power supply input specifications, and is
set in the SQM-242 setup. Max Power controls the maximum % of the full scale power
that can be used by this process layer.

Slew Rate: The maximum power change allowed on an output, per second. If rate
ramps or PID power requirements exceed this value, an error will occur.

Source Index: If using a source index, assigns each film to a specific pocket (one of 16
values). These values are sent to the PLC at the beginning of each layer. See the
System Setup for other settings that pertain to the source index.

Output: Assigns the selected material to one of the source control outputs. When an
output is selected, the Sensor Tooling panel changes to show only those sensors that
are assigned to that output in the Edit, System dialog. The combination of a source
output and its sensor inputs defines the deposition “control loop” for the selected
material.