KD Scientific Allego Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

Tube Calibration

Please use the following procedure to calibrate tubing.

To achieve the best possible accuracy, the calibration

procedure should be run before using the pump. Please

note that the pumping direction for calibration is always

forward.

1. From the Configuration screen, make

appropriate selections for tubing size and flow

rate and select the Calibrate button and the

Calibration screen will open.

2. Insert appropriate tubing into Pump Head.

Please see the previous section for tubing

break-in procedure.

3. Insert tube inlet into supply fluid.

4. Press and hold the fast forward button to

prime the tubing. Priming will stop when the

button is released.

5. Insert tube outlet into a container to be placed

on a scale. The weight to volume conversion

for water is 1 gram = 1 ml.

6. Press the Run button. The drive will run until

stopped by the user.

Select Calibrate when tubing and

flow rate have been entered

Select Fast Forward

to prime the tubing

Select Run

to start the

calibration

program