KD Scientific Allegro Peristaltic Pump Series User Manual

Page 51

Advertising
background image

51

K

D

S

A

lle

g

ro

P

e

ri

s

ta

lt

ic

P

u

m
p

S

y

s

te

m

U

s

e

r’s

M
a

n

u

a

l

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

KD Scientific Allegro Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

Operating the KDS Allegro Peristaltic Pump System

I/O (IN) Event

The I/O (IN) Event command is used to have the pump

pause and wait for a signal from an external device or

for a manual user input at the touchscreen. These

external signals are communicated via Pin 2 on the

15-pin I/O Connector port.

The following options are available when setting an I/O

(IN) Event command.

Input High-to-Low: Sets the pump to wait for

a logic low input signal falling edge.

Input Low-to-High: Sets the pump to wait for

a logic high input signal rising edge.

User Press: Sets the pump to wait for manual

input from the user at the touchscreen

(screen touch).

TO INSERT AN I/O (IN) EVENT COMMAND:

1.

From the Program Main screen, choose

Step Definition to enter the Step Definition screen.

2. Highlight (select once) the Program Step that you

want the I/O (IN) Event to follow.

3. Choose the

button to access the

Program Steps Menu screen.

4. Choose

Advanced Options

to enter the Select

Advanced Options screen.

5. Choose the

I/O (IN) Event

button to enter the

Event Step Definition Screen.

6. Select either Input High, Input Low or User Press,

as appropriate.

7.

Choose the Accept button to save your changes

and return to the Step Definition screen.

‘INS STEP’

I/O (IN) Event Advanced Options menu

Advertising