Configuring the KDS Allegro Peristaltic Pump System

K

D

S

A

lle

g

ro

P

e

ri

s

ta

lt

ic

P

u

m
p

S

y

s

te

m

U

s

e

r’s

M
a

n

u

a

l

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

KD Scientific Allego Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

7. Upon completion of the calibration run period,

the ‘Enter Volume Delivered’ button will be

activated. Please note the Volume reading listed

under the Tubing and Rate information on the

Calibration screen.

8. Select the Enter Volume Delivered button and

enter the Volume as measured on your scale

into the keypad and press Enter.

9. Press Enter on the Calibration screen to save

the setting or select the Run button to start the

procedure over again, or press the Reset

Calibration button to reset to the Factory

default value.