KD Scientific Allego Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

TO SET UP/EDIT A PULSE RATE PROFILE:

1.

From the Pulse

Definition

screen for a Pulse Rate

profile, choose the Direction button to toggle the

Flow Mode between Infuse and Withdraw

operation.

2. Choose the

Rates

button to enter the Set Pulse

Rates screen.

3. Select the

Rate 1

button (it should be selected by

default as indicated by the active data entry box to

the right of the button). Enter the flow rate value

desired using the keypad or select the

Minimum

Rate or Maximum

Rate button.

4. If the Units of Rate are not correct for your

application, choose the

Units

button.

5. On the Units Selection screen, select the unit of

measure (L, ml, µl or nl) and the unit of time (sec,

min, hr) as appropriate, then choose Accept to

confirm your selections.

6. Repeat Steps 3 to 5 for the second flow rate by

selecting the

Rate 2

button and continuing

data entry.

7.

When all parameters are correctly set on the Set

Pulse Rates screen, choose Accept to confirm your

settings and return to Pulse Definition screen.

8. Select the Volume or Time button with indicator to

determine how the pump cycle length will be

managed. Volume calculates each cycle stage by

pumping for the length of time needed to pump the

specified volume. Time manages each cycle stage

by pumping for the amount of time specified.

9. Depending on whether you have selected

the

Volume

or

Time

button in the preceding step,

choose the

Time

or the

Volume

button to enter the

Set Pulse Times or the Set Pulse Volumes screens.

Set Pulse Times screen used to define the length of time

the pump will operate at each pulse step level

Step Definition screen for a Pulse Profile

with Volume mode selected

10. If you selected Set Time, continue with Step 11.

If you selected Set Volume, proceed to Step 14.

11. On the Set Pulse Times screen, select the

Time 1

button (it should be selected by default as

indicated by the active data entry box to the right

of the button). Enter the desired value using the

keypad. Repeat for Time 2.

12. If the Units of Time associated with Time 1 or Time

2 are not correct for your application, select the

Units button associated with either parameter to

toggle between sec (seconds) and h:m:s

(hours:minutes:seconds). To use the h:m:s option,

enter your time value in hours, minutes and

seconds, separating each value with a colon.

13. When both Time parameters are set to the correct

values, choose the Accept button to return to the

Pulse Definition screen. Continue with Step 17.

14. On the Set Pulse Volumes screen, select the Volume

1 button (it should be selected by default as

indicated by the active data entry box to the right

of the button). Enter the desired value using the

keypad. Repeat for Volume 2.

15. If the Units of Volume associated with

Volume 1

or

Volume 2 are not correct for your application,

choose the

Units

button associated with either

parameter to enter the Units selection screen.

Select the unit of measure button (ml, µl, nl, pl)

applicable to your application, then choose the

Accept button to return to the Set Pulse Volumes

screen.

16. When both Volume parameters are set to the

correct values, choose the Accept button to return

to the Pulse Definition screen.

17. Select the

# of Pulses

button to enter the Set #

Pulses screen.

18. On the Set # Pulses screen, enter the desired value

using the keypad, then choose the Accept button.