Upgrading Allegro Peristaltic Pump System Software

1. Upload the latest software version to your desktop (format is filename.srec).

2. Disconnect all I/O devices and then connect the pump to PC using a USB cable.

3. On the Set up screen of the pump, press the Upgrade Software button twice to enter “Boot Loader

Mode”.

4. From the CD provided with the pump, open the Allegro Peristaltic Pump System Pump Updater

Application.

5. Click Open Update File and browse to the file previously saved on your desktop and click Open.

6. Click Start Update.

7. After the update is complete, press Reboot to power cycle the pump.

Appendices

NOTE: Prior to upgrading, users will need to

install the KDS Allegro Peristaltic Pump System

USB driver as well as the bootloader driver

file (KDS Allegro Peristaltic Pump System-

Bootloader.inf). Reference “USB Virtual

Comport Driver Installation”.

NOTE: To exit boot loader mode without

upgrading the software, power cycle the

pump.