APPENDIX C: EXTERNAL CONNECTIONS

User I/O Connector Specifications

Pin Assignments

1.

Direction control input

Rising edge sets pump to forward

Falling edge sets pump to reverse

2.

Trigger Input

Event trigger – falling edge triggers a program event

3.

Footswitch Input (or Timer)

Settable to work as:

-Momentary switch closure to ground or TTL Logic

Low; falling edge toggles between states (run/stop)

-Starts on rising edge, stops on falling edge

-Starts on falling edge, stops on rising edge

4.

Trigger 1 Output

TTL Logic Output-Default = Low Control thru program

or serial comm

5.

Trigger 2 Output

TTL Logic Output-Default = Low Control thru program

or serial comm

6.

Sync Output

TTL Logic Output –

Rising Edge = Start Forward

Falling Edge= Start Reverse

7.

Direction Output

TTL Logic Output –

High = Reverse

Low = Forward

8.

Valve Output (controls voltage only)

TTL Logic Output –

High = Valve Actuated

Low = Valve Off

Note: Valve follows motor direction,

when valve is set to Auto.

9-13. Signal Return / Ground

14.

Run Indicator Voltage

+5V through a 470 ohm resistor (connect to LED

Anode)

15.

Run Indicator Output

TTL Logic Output, active low (low = run) (connect to

LED Cathode)

Electrical Specifications

Inputs:

V

IH

≥ 2V

I

IH

≤ 20µA

V

IL

≤ 0.4V

I

IL

≤ 0.5mA

Outputs:

V

OH

≥ 3.8V

I

OH

≤ 6mA

V

OL

≤ 0.4V

I

OL

≤ 6mA

All Inputs are pulled high to +5 Vdc through a 10k

resistor All Outputs are at TTL Logic Levels. Pulse duration

should be 0.1s minimum.

User I/O Connector Specifications