KD Scientific Allegro Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

Operating the KDS Allegro Peristaltic Pump System

ALLEGRO PERISTALTIC PUMP SYSTEM

PROGRAM EXPORT AND IMPORT

The Allegro Peristaltic Pump System pump permits the

daisy chaining of multiple pumps, as well as integration

with/control by an external computer through USB. To

facilitiate operation in these modes, the Allegro Peristaltic

Pump System application software includes a variety of

commands designed to simplify the export/import of

Programs between the pump and external devices. This

section outlines the use of these commands.

TRANSFER A PROGRAM BET WEEN PUMPS

Follow these instructions to transfer a Program between

two KDS Allegro Peristaltic Pump System pumps.

1.

Connect two pumps together via an RS-485

(firewire type) cable and set the address on the first

pump to ‘00’ using the Set Pump Address

command in the Settings menu.

2. Set the address on the second pump to something

other than ‘00’ using the Set Pump Address

command in the Settings menu.

3. From the Program Main screen, choose Program

Select then choose the Program you would like

to export.

4. Choose the

File Options

button.

5. Choose the Pump button and enter the address for

the pump you would like to receive the Program.

6. Choose Export Program to Pump.

7.

The pump will display a confirmation message in

the Message Area on the screen when the export

is complete. Note that long Programs may take a

few seconds to export.

8. Before disconnecting the pumps, verify that the

Program was successfully exported by choosing

Program Select from the Program Main screen on

the destination pump, then browse the available

Programs to verify that the exported Program is

shown in the menu.

NOTE: Do not attempt to Export/Import

Programs while the pump is running.

NOTE: Do not change screens while

Importing/Exporting a program.

NOTE: For additional details on the

physical interfaces used to connect the KDS

Allegro Peristaltic Pump System to external

devices, refer to the External I/O

Connections section in the Appendices.

Export Program confirmation screen showing successful export

EXTERNAL PUMP CONTROL

The Allegro Peristaltic Pump System can be controlled by

a terminal communication program. KD Scientific

provides a Pump Control program on the CD that came

with the unit. The Program Manager software and user

manual can be found in the Pump Applications directory

on this CD.