KD Scientific Allegro Peristaltic Pump Series User Manual

Page 42

Advertising
background image

42

K

D

S

A

lle

g

ro

P

e

ri

s

ta

lt

ic

P

u

m
p

S

y

s

te

m

U

s

e

r’s

M
a

n

u

a

l

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

KD Scientific Allegro Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

Operating the KDS Allegro Peristaltic Pump System

Flow

Rate

Time

STOP

Target Time

Ramp Profile

The Ramp Rate profile is used to create a linearly increasing

or decreasing flow rate for a single pumping mode (infuse

or withdraw) over the course of a defined time period.

Ramp Rate Pump Operation Profile

When creating or editing a Ramp Rate profile, you will

define three primary operating parameters:

The Flow Mode defining the direction of flow:

forward or reverse

The Flow Rate at the beginning and end of

the pumping cycle

The Target Time that determines the total length

of time over which the ramp will be calculated

During a Ramp Rate profile, the KDS Allegro Peristaltic

Pump System varies the flow rate linearly between the

starting rate and ending rate over the full time period

entered. A Ramp Rate profile can be applied to an

increasing or decreasing rate application.

When creating or editing a Ramp Rate profile, you can set

the rate in L, ml, µl, nl or pl units of volume using sec or

h:m:s (hours:minutes:seconds) as the units of time. The

system automatically calculates a minimum and maximum

flow rate based on the tubing selected or custom-defined.

The min/max values are made available as button selection

options on the Rate screen.

TO SET UP/EDIT A RAMP RATE PROFILE:

1.

From the Step Definition screen for a Ramp Rate

profile, choose the Direction button to toggle the

Flow Mode between Forward and Reverse operation.

2. Choose the Set

Start & End Rate

button to enter

the Set Start & End Rates screen.

3. Select the

Start Rate

button (it should be selected

by default as indicated by the active data entry

box to the right of the button). Enter the flow rate

value desired using the keypad or select the

Minimum Rate

or

Maximum Rate

button.

4. If the Units of Rate are not correct for your

application, choose the

Units

button.

5. On the Units Selection screen, select the unit

of measure (L, ml, µl or nl) and the unit of time

(sec, min, hr) as appropriate, then

choose

to confirm your selections.

6. Repeat Steps 3 to 5 for the End Rate by selecting

the

End Rate

button and continuing data entry.

7.

When all parameters are correctly set on the

Set Rate screen, choose

to confirm your

settings and return to Ramp Definition screen.

8. Choose the Time button to enter the Set Time screen.

9. Enter the desired Time value using the keypad.

10. If the Units of Time are not correct for your

application, select the

Units

button to toggle

between sec (seconds) and h:m:s

(hours:minutes:seconds). To use the h:m:s option,

enter your time value in hours, minutes and

seconds, separating each value with a colon.

11. When all parameters are correctly set on the Set

Target Time screen, choose Accept to confirm your

settings and return to the Ramp Definition screen.

12. Choose Accept again to return to the Step

Definition screen.

‘ENTR’

‘ENTR’

Step Definition screen for a Ramp Rate profile

Set Start and End Rates screen

NOTE: Ramps < 2 seconds from min to

max rates may cause motor stalls.

Advertising