Operating the KDS Allegro Peristaltic Pump System

Flow

Rate

Time | Volume

STOP

Target Time or

Total Volume

Point

Constant Rate Profile

The Constant Rate profile is used to run the pump at a

defined flow rate and for a specific pumping mode

(infuse or withdraw) up to a target volume or for a

defined time period.

Constant Rate Pump Operation Profile

When creating or editing a Constant Rate profile, you

will define three primary operating parameters:

The Flow Mode defining the direction of flow:

forward or reverse

The Flow Rate to define the rate of uptake or

dispensing

The Target Volume or Run Time that determine how

long the pump will operate at the set rate

When creating or editing a Constant Rate profile, you

can set the rate in L, ml, µl or nl units of volume using sec,

min or hr as the units of time. The system automatically

calculates a minimum and maximum flow rate based on

the tubing size selected or custom-defined. The min/max

values are made available as button selection options on

the Rate screen.

TO SET UP/EDIT A CONSTANT RATE PROFILE:

1.

From the Step Definition screen for a Constant Rate

profile, choose the Direction button to toggle the

Flow Mode between Infuse and Withdraw operation.

2. Choose the

Rate

button to enter the Set Rate screen.

3. Enter the flow rate value desired using the keypad

or select the

Minimum Rate

or

Maximum Rate

button.

4. If the Units of Rate are not correct for your

application, choose the

Units

button.

5. On the Units Selection screen, select the unit of

measure (L, ml, µl, nl or pl) and the unit of time

(sec, min, hr) as appropriate, then choose

to confirm your selections.

6. When all parameters are correctly set on the Set

Rate screen, choose

to confirm your

settings and return to Constant Rate Definition screen.

7.

Choose the

Volume

button to enter the Set Volume

or Time screen.

8. On the Set Volume or Time screen, select the

Volume or Time

button, as desired.

9. Enter the desired value using the keypad, or select

the

Maximum Volume

button.

10. If the Units of Volume/Time are not correct for your

application, choose the

Units

button to the right of

the parameter entry box. For the Units of Volume,

this choice takes you to the Units Selection screen.

For the Units of Time option, this choice will toggle

the button between sec (seconds) and h:m:s

(hours:minutes:seconds). To use the h:m:s option,

enter your time value in hours, minutes and

seconds, separating each value with a colon.

11. If you are changing the Units of Volume, on the

Units Selection screen, select the unit of measure

(ml, µl, nl or pl) as appropriate, then choose

to confirm your selection.

12. When all parameters are correctly set on the Set

Target Volume or Time screen, choose

to

confirm your settings and return to the Constant

Rate Definition screen.

13. Choose

again to return to the

Step Definition screen.

‘ENTR’

‘ENTR’

‘ENTR’

‘ENTR’

‘ENTR’

Constant Rate Profile Setup Screen