KD Scientific Allegro Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

Appendices

RS-485

1

2

3

4

5

6

Pin #

RS-485 In Signal

RS-485 Out Signal

Notes

1

PWR IN

PWR OUT (+30V)

*

2

GND

GND

3

RS-485 A

TIED TO PIN 4

**

4

RS-485 B

TIED TO PIN 3

**

5

TIED TO PIN 6

**

RS-485 A

6

TIED TO PIN 5

**

RS-485 B

Mating Face

*

Protected by

resettable fuse

**

100 Ω, ½ w to

GND termination

RS-485 In/Out - IEEE-1394 Sockets

Daisy-Chaining via RS-485

P.C.

PUMP

ADDR Ø

USB

USB

RS-485

OUT

PUMP

ADDR 1

RS-485

OUT

RS-485

IN

PUMP

ADDR n

RS-485

IN

RS-485

IEEE 1394

Cable

RS-485

• • •

IEEE 1394

Cable

!

CAUTION: DO NOT CONNECT TO FIREWIRE PORTS ON A PC. DAMAGE MAY
OCCUR TO PUMP AND/OR PC.

NOTE: When applying power to pumps in a daisy chain, apply power to the last group in the chain first

and work backwards.

NOTE: When removing power, be aware that power removed from a stand-alone unit will cause power to

be removed from all RS-485 powered units deriving power from that stand-alone unit.

NOTE: Power may be removed from any group of pumps, except the first group, without affecting

communications “downstream”

NOTE: Each pump in the daisy-chain must have a unique address. The first pump must be set to address 00.