KD Scientific Allegro Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

Appendices

TROUBLESHOOTING

1. Infusion Accuracy

To ensure infusion accuracy always use new tubing and

measure tubing bore diameter and enter actual

dimensions in millimeters (mm) using the Custom Tubing

entry option. Additionally, make sure that the guide rods

and lead screw are properly lubricated.

NOTE: It is recommended that you save

your programs periodically by uploading

them to a PC.

KDS Allegro Peristaltic Pump System

Display Error Messages

2. Motor Stalled

This indicates that the pusher block travel has been

impeded. This may be caused by the tubing plunger

hitting bottom, a kink in the tubing (occlusion), tubing

plunger binding or any situation requiring more force to

the head of the tubing than the pump is capable of

delivering. To restart the pump, remove the cause of

stalling and/or increase the force setting, then press

RUN. The pump will continue the interrupted procedure

where it stopped.

3. Out of Range

A value was entered or encountered in a pump program

that was beyond the pump’s limits.

4. Power Failure Notification

If power to the pump is interrupted during operation, an

alarm will sound to alert the user to the interruption.

When power is restored, an onscreen message is

displayed to indicate the interruption.

5. Unit Won’t Power Up

Verify the the Power Source Switch on the rear of the

pump is set to DC-IN if powering through power supply

or RS-485 if powering in a chain.

The KDS Allegro Peristaltic Pump System Pump will notify you of a power

failure with a message when power is restored