Pulse Profile

The Pulse Rate profile is used to create an oscillating

infusion or withdrawal profile that varies between a

starting flow rate and a second rate that can be either or

higher or lower than the start rate. The flow rate is

constant within each segment of the pulsed flow profile

and the user defines the total number of repetitions

(pulses) for the oscillating pattern. Each segment of the

profile must be set to run for either a specific length of

time or to deliver a set volume.

Volume

Time | Volume

# Pulses

Pulse Rate Pump Operation Profile

When creating or editing a Pulse Rate profile, you will

define four primary operating parameters:

The Flow Mode defining the direction of flow:

forward or reverse

The Flow Rate to be used at each level of the

pulsed profile within the pumping cycle

The length of Time that each segment of the profile

will run or the specific Volume to be

infused/withdrawn during each segment of the

profile

The Number of Pulses (repetitions) that will be

executed during the pumping cycle

During a Pulse Rate profile, the KDS Allegro Peristaltic

Pump calculates a series of equal length pumping stages

at two user-defined flow rates: a starting rate and a

“pulse” rate. You enter the Number of Pulses (repetitions)

that you want to use, along with how you want the system

to calculate the length of each stage — by time or by

volume pumped. A Pulse Rate profile can be applied to

pulses that are higher in rate or lower in rate than the

starting flow rate.

When creating or editing a Pulse Rate profile, you can set

the rate in L, ml, µl or nl units of volume using sec, min or

hr as the units of time. The system automatically

calculates a minimum and maximum flow rate based on

the syringe selected or custom-defined. The min/max

values are made available as button selection options on

the Set Pulse Rates screens.