KD Scientific Allego Peristaltic Pump System User’s Manual

Publication 5613-004-REV-A

The break in programs are installed on the pump.

They are also located on the CD that ships with the pump.

KD Scientific may change the programs so are shown here

for reference only.

Allegro I: “I_BREAK_IN”

1. Ramp

minimum flow to 300 ml/min

2½ min.

2. Ramp

300 to 650 ml/min

2½ min.

3. Constant

700 ml/min

5 min.

4. Constant

800 ml/min

5 min.

Allegro II: “II_BREAK_IN”

1. Stepped

→5

minimum to 10 µl/min

1 min.

2. Stepped

←5

10 to 50 µl/min

2 min.

3. Ramp

50 µl/min to 10 ml/min

3 min.

4. Ramp

10 to 200 ml/min

3 min.

5. Stepped

→15 200 to 800 ml/min

5 min.

6. Constant

850 ml/min

3 min.

7. Constant

1700 ml/min

3 min.

Allegro III: “III_BREAK_IN”

1. Ramp

minimum to 100 ml/min

2½ min

2. Ramp

100 to 250 ml/min

2½ min.

3. Stepped

→15 500 to 1000 ml/min

2½ min.

4. Stepped

←15 900 to 1500 ml/min

2½ min.

5. Constant

1700 ml/min

5 min.

6. Constant

1700 ml/min

5 min.

APPENDIX B: TUBE BREAK-IN PROGRAMS