3 safety devices, Safety devices, Safety – Leica Biosystems CM1860/CM1860 UV User Manual

Page 10

Advertising
background image

10

Instructions for Use V1.4 – 10/2012

2. Safety

UVC disinfection is effective when disinfecting sur-

faces and air within the irradiated working area of the

Leica CM1850UV, CM1860UV and CM1950 cryostats at -20 °C

( Table 1, see I. Maier certificate). For powerful disinfection, we

recommend irradiation for three hours (CM1850UV/CM1860UV/

CM1950). Vegetative bacteria including Mycobacterium tuber-

culosis, bacterial endospores (Bacillus sp.) and fungi are killed

during this time. Even viruses including such resistant species

as hepatitis are inactivated up to at least 4 log10 units (99.99 %).

2.3 Safety devices

To prevent adverse health effects from UV radiation, the UV dis-

infection cycle can be started only after the sliding window has

been properly closed. Closing the window activates the corre-

sponding safety features.

When working with cryostats, personal safety precautions must

always be taken. It is mandatory to wear work safety shoes, safe-

ty gloves, a mask and safety goggles.

The consistent use of these safety features and strict observation of the

warnings and cautions in these Instructions for Use will safeguard the

operator from accidents and/or personal injury to a great extent.

Personal safety precautions

The instrument is equipped with the following safety devices: Handwheel

lock and knife guard on the knife holder.