Leica CM1860/CM1860UV – Cryostat

To start disinfection, the sliding window must be completely closed.
Press the UVC button once briefly, 30 min. mode

Press the UVC button once for a longer time (approx. 4sec) to start the 180 min. mode

Both LEDs on the display for short and long disinfection indicate the status of the disinfection process:

Status

LED

LED

UVC lamp

Short disinfection

Long disinfection

Ready

1

OFF OFF OFF

Service life

alternating flashing

OFF

expired

2

Short disinfection

ON

OFF

ON

active
Long disinfection

OFF

ON

ON

active
Short disinfection

Flashing OFF

OFF

interrupted

3

long disinfection

OFF

Flashing OFF

interrupted

4

1

Status: Ready

Both LEDs and the UVC lamp are off.
• Start short disinfection: Press the UVC button briefly (<1 second). The LED for the short disinfection

cycle will illuminate.

• Start long disinfection: Press the UVC button for longer (>1 second). The LED for the short disinfec-

tion goes out and the LED for the long disinfection goes on.

2

Status: Service life of the UVC lamp expired

The LEDs for the short and long disinfection modes illuminate alternately, and the UVC lamp is off.
This status is displayed as soon as the service life of the UVC lamp (9000 hours) has been reached.

For replacing the UVC lamp, refer to Chapter 9.3.3.