background image

23

Leica CM1860/CM1860UV – Cryostat

4.4 Assembling the handwheel

The handwheel and all parts for installation are included in the carton with the accessories.

The handwheel can be dismounted for transporting (e.g. narrow doors).

To install the handwheel, follow these steps:
• Insert the pin (1) of the handwheel shaft into the hole (2) of the handwheel.
• Attach the spring washer (3) to the screw (4) as shown in Fig. 10.
• Tighten the screw (4) using an Allen key (6 mm).
• Attach the cover disk (self-adhesive - not pictured).
To dismount, proceed in reverse order.

4. Setting up the Instrument

Rotate the handwheel only if the refrigeration system is on and the cryochamber is cold.