Cleaning, disinfection, maintenance – Leica Biosystems CM1860/CM1860 UV User Manual

Page 67

Advertising
background image

67

Leica CM1860/CM1860UV – Cryostat

Removal of the lamp
The UVC lamp (3) is installed in front

of the protection screen for chamber

illumination (4).
• Hold the lamp with both hands and

carefully pull it out of the clips (5)

with a slight forward movement.

• Detach the metal ring (7) on the

holder (6) in the direction of the ar-

row (8) and carefully pull the lamp

out of the holder with your right

hand (

see Fig. 60 + 61

).

Installation of the new lamp
• Carefully slide the metal ring (7) over

the lamp from the left (

see Fig. 62

).

• Push the lamp into the holder on the

left side until it engages.

• Screw the metal ring onto the

holder, then hold the lamp with both

hands and carefully push it into the

clips (5).

• Replace the sliding window.
• Connect the instrument to the power

supply again and switch it on.

4

3

5

7

Fig. 60

Fig. 61

Fig. 62

6

8

Dispose of the UVC lamp separately!

9. Cleaning, Disinfection, Maintenance

If any metallic mercury is

released, handle it carefully

and dispose of it properly!