Leica CM1860/CM1860UV – Cryostat

7. Daily Use of the Instrument

7.3 Sectioning
7.3.1 Blade holder CE

1

2

3

When inserting the blade/knife holder

base, you have to overcome the resis-

tance of a spring (located in the foot

of the knife holder) by applying a little

pressure towards the left.

Inserting the base

1. To insert the blade/knife holder base (1), move

the lever (2) to the back. Slide the base onto

the T-piece (3) of the baseplate.

2. To clamp the blade/knife holder base, turn the

lever (2) to the front.

Inserting the blade holder CE

3. Push the upper part of the blade holder onto

the base. Clamp the blade holder into place

using an Allen key (No. 4) on the left side (23).

Fig. 27

23

Fig. 26

If an additional blade holder instruction manual has been provided,

read and follow the instructions and safety information carefully!