Daily use of the instrument – Leica Biosystems CM1860/CM1860 UV User Manual

Page 49

Advertising
background image

49

Leica CM1860/CM1860UV – Cryostat

Blade holder CE

1. To release the segment arc (2) from the base,

use a 4 mm Allen key to loosen the screw

over the clearance angle adjustment (23) and

remove the segment arc (2) from the blade/

knife holder base.

2. Fold the anti-roll guide system (4) to the left

while holding the lever (11).

3. Open the clamping lever (10,

Fig. 49

) by turning

it counterclockwise and pull it out.

4. Afterwards, the pressure plate (36) can be

removed for cleaning (with alcohol).

5. Open the clamping lever of the blade holder

(12,

Fig. 49

) by turning it counterclockwise and

pull it out. The blade holder can now be moved

sideways and can be taken out of the segment

arc.

36

4

11

Fig. 49

12

Knife holder CN

Fig. 50

Occasionally oil the moving parts such as the

axes (37) and (38) and the slot (39) with a drop of

cryostat oil (Type 407).

39

38

37

Disinfection

Also observe the warning message on

Page 13!

Place all of the removed parts into a disinfect-

ing solution for the time recommended by the

manufacturer. Dry each part thoroughly before

reassembly and insertion into the cryostat.

7.

Daily Use of the Instrument

10

23

2

If several blade/knife holders are

cleaned at the same time, the parts

must NOT be mixed up! Failure to ad-

here to this may result in sectioning

problems!