Instructions for Use V1.4 – 10/2012

7.

Daily Use of the Instrument

Fig. 40

7.3.3 Knife holder CN

1

2

3

When inserting the blade/knife holder

base, you have to overcome the resis-

tance of a spring (located in the foot

of the knife holder) by applying a little

pressure towards the left.

Inserting the base

1. To insert the blade/knife holder base (1), move

the lever (2) to the back. Slide the base onto

the T-piece (3) of the baseplate.

2. To clamp the blade/knife holder base, turn the

lever (2) to the front.

Inserting knife holder CN

3. Push the upper part of the knife holder (3) onto

the base (1). Clamp the knife holder into place

using an Allen key (No. 4) on the left side (23).

Fig. 41

1

3

23

Clearance angle settings of 4° - 6° (knife

holder CN and blade holder CE-TC) are

well suited for most applications.