2 blade holder ce-tc, Daily use of the instrument – Leica Biosystems CM1860/CM1860 UV User Manual

Page 43

Advertising
background image

43

Leica CM1860/CM1860UV – Cryostat

7. Daily Use of the Instrument

10

9

8

6

7

Assembling the anti-roll guide system/replacing the anti-roll plate

1. Insert the glass into the interchangeable frame

and tighten it evenly using the knurled screws (7).

2. Insert the shaft (8) of the metal frame for ex-

changeable glass inserts from above into the

bore of the swinging arm (6) in such a way that

the pin rests in the notch.

3. Push the white plastic plate (9) from below

onto the shaft (8).

4. Screw the knurled nut (10) from below onto

the shaft (8).

All 4 lengths of the anti-roll guide glass

can be used.

7.3.2 Blade holder CE-TC

The blade holder CE-TC is designed only for dis-

posable carbide metal blades (tungsten carbide

- TC65).

The handling of this holder is identical to that of

the blade holder CE.

Fig. 38

Fig. 39