Cleaning, Disinfection, Maintenance

9.3.1 General maintenance instructions

The microtome is virtually maintenance-free. To

ensure a smooth operation of the instrument over

several years we recommend the following:

• At least once a year, have the instrument

inspected by a qualified service engineer

authorized by Leica.

• Enter into a service contract at the end of

the warranty period. For further information,

please contact your local Leica service center.

• Clean the instrument every day.

Weekly:

Before the following parts can be

oiled, all section waste and deposits

must be carefully removed.

• Oil the plastic coupling (6,

Detail 57a

) with a

drop of cryostat oil (Type 407);

• Lubricate the specimen cylinder (1):

Push the appropriate coarse feed button to

move the specimen cylinder out to the front

stop position, apply a drop of cryostat oil and

move the specimen cylinder back to the home

position by pressing the appropriate coarse

feed button.

Occasionally, or when required:

• Apply a drop of cryostat oil (Type 407) to the

clamping piece (T-piece) (2,

Fig. 55

) on the mi-

crotome base plate and the clamping lever (3);

9.3 Maintenance