Operating the instrument – Leica Biosystems CM1860/CM1860 UV User Manual

Page 30

Advertising
background image

30

Instructions for Use V1.4 – 10/2012

After replacing the UVC lamp, its service life counter must be reset. To do so, press and hold the

UVC button for longer than 30 seconds. The counter has been reset successfully if the LED for the

long disinfection goes out while the button is being pressed. Once the UVC button is released,

the disinfection status goes into ready mode.

3

Status: Short disinfection interrupted

The LED for short disinfection flashes, and the UVC lamp is off. The operator must press the UVC button

to confirm. The disinfection system goes into ready mode, or, if the service life of the UVC lamp has

expired, both LEDs flash in alternately.

4

Status: Long disinfection interrupted

The LED for long disinfection flashes, and the UVC lamp is off. The operator must press the UVC button

to confirm. The disinfection system goes into ready mode, or, if the service life of the UVC lamp has

expired, both LEDs flash alternately.

The disinfection mode is interrupted as soon as the sliding window of the cryostat is open or the

operator presses the UVC button. In the case of a power failure, when the instrument restarts it

checks whether it was running a disinfection cycle, and if so which one. The LED of the determined

disinfection cycle (short or long disinfection) flashes.

6.

Operating the Instrument