7.5.3 Manual defrosting of the quick freeze shelf

The quick freeze shelf may become hot during defrosting! There-

fore, do not touch it!

In case of heavy frost formation on the quick freeze shelf, which can occur

particularly after the spray disinfection, a defrost can be activated manually

(

see point 6.2.5

). which can be terminated when required.

7.6 Terminating work
7.6.1 Terminating daily work

• Lock the handwheel.
• Take the knife out of the knife holder and put it back into the knife box in

the cryochamber.

• Remove frozen section waste with a cold brush.
• Empty the section waste tray.
• Clean the storage shelves and brush shelf.
• Remove all tools and supplies that would prevent the UV light from reach-

ing the chamber walls and shelves.

To clean and disinfect the interior of the instrument, use only alco-

hol-based disinfectants. On the exterior, common detergents and

alcohol-based disinfectants can be used.

All components removed from the cold environment will collect

condensation. Therefore, dry them thoroughly before placing them

back into the cryochamber.

• Remove all the specimens from the cryostat.

• Attach the cover for the freeze shelf (standard scope of delivery).

• Remove the plug and close the sliding window.

• Switch off the chamber illumination and start the UV disinfection.

• Lock control panel field 1 (

Fig. 14

) using the key button.

• Do NOT switch off the instrument using the circuit breaker, as otherwise

no cooling takes place.

7. Daily Use of the Instrument

Be sure to place the corresponding cover on the quick freeze shelf

to prevent frost. Always cover the quick freeze shelf during breaks

and overnight.