7.

Daily Use of the Instrument

Fig. 34

The safety gloves included in the stan-

dard scope of delivery must be worn

when disposing of the blade!

5. Once the blade has been removed from the

blade holder, it is disposed of into the dis-

penser container (storage compartment in

bottom, Fig. 34).

Lateral movement

It is only necessary to clamp a disposable blade

into the CE holder one time. When a portion of the

blade becomes dull or nicked, the upper portion

of the blade holder (shown on its base in Fig. 35)

will move from side to side to allow the use of the

entire blade edge. When moved, the alignment of

block to blade will be very close to the previous

position since the clamping pressure on the blade

remains the same.
To reposition the blade holder, follow these steps:
1. Turn the clamping lever (14) to the rear (clock-

wise) to release, and then shift the blade

holder sideways to the desired position.

2. To tighten, turn the clamping lever (14) to the

front (counterclockwise).