Leica CM1860/CM1860UV – Cryostat

Medium disinfection can be attained through short irradiation for

30 minutes (CM1850UV/CM1860UV/CM1950). This reduces vegeta-

tive bacteria including Mycobacterium tuberculosis and sensitive

viruses such as the influenza A virus (also including the highly

pathogenic avian influenza virus H5N1) and the poliovirus by at

least 5 Log 10 units (99.999 %). UVC irradiation within the working

area of the cryostats can provide reliable and efficient disinfection

of surfaces and the air and significantly reduces the risk of infec-

tion.
We recommend removing tools and debris and then wiping off vis-

ible contamination in the cryostat with an alcohol-based disinfec-

tant prior to using the UV lamps. The germicidal effect of the irra-

diation is restricted to the directly irradiated areas, which is why

UVC irradiation cannot be a replacement for regular chemical dis-

infection of the cryostat chamber.

For current information about certificates and recommendations,

visit:
http://www.LeicaBiosystems.com/products/total-histology/cryo-

sectioning/details/product/leica-CM1860-uv/downloads/