Instructions for Use V1.4 – 10/2012

7.

Daily Use of the Instrument

Inserting the blades into blade holder CE

The blade holder CE can be used either for the low-profile, narrow-band and high-profile,

broad-band blades.

Caution!

Microtome blades are extremely sharp!

10

9

4

11

Inserting the high-profile blades

Fig. 29

The safety gloves included in the stan-

dard scope of delivery must be worn

when inserting the blade!

The anti-roll guide system functions as

a knife guard here!

1. Fold the anti-roll guide system (4) to the left – while

doing so, hold the lever (11) (not the adjusting

screw of the anti-roll guide) so that the height of

the anti-roll guide remains unchanged.

2. Open the clamping lever (10) by turning it

counterclockwise (

Fig. 28

).

3. Carefully insert the blade (9) from above or

from the side between the pressure plate and

the blade rest. Make sure that the blade is

inserted so that it is centered and lies evenly

along the ledge (

see red arrow in Fig. 28

).

4. Clamp the clamping lever (10) by turning it

clockwise (

see Fig. 29

).

5. Fold the anti-roll guide system (4) back to the

right (toward the blade) using the lever (11).

Fig. 28

Inserting low-profile blades

• When using low-profile blades, the red guide

(blade rest) must first be placed in the blade

holder, and then the blade.

Fig. 30

4

11

10

Advertising