7.6.2 Shutdown for a longer period

If you do not intend to use the instrument for several weeks you may

turn it off.

Please note, however, that it may take up to several hours to cool

the cryochamber down to very low temperatures after turning the

instrument on again.

After turning off, the instrument should be cleaned and disinfected

thoroughly (see chapter "9. Cleaning, Disinfection, Maintenance").

• Follow the steps in 7.6.1 until UV disinfection is complete.
• Switch the instrument off at the circuit breaker and unplug the power

plug.

• Open the sliding window to allow air circulation, allow the chamber to

dry and prevent accidental breakage while stored.

Switching off the instrument via the circuit breaker will not affect the pro-

grammed values.
Before turning the instrument on again, the cryochamber microtome and all

accessory components must be absolutely dry.