Technical data – Leica Biosystems CM1860/CM1860 UV User Manual

Page 16

Advertising
background image

16

Instructions for Use V1.4 – 10/2012

3.

Technical Data

Operating temperature range (ambient temperature): 18 °C to 35 °C. All specifications related to

temperature of the cooling unit are valid only for an ambient temperature of 22 °C and a relative

humidity of no more than 60 %!

Type

CM1860/ CM1860/ CM1860/ CM1860/ CM1860/

CM1860 UV

CM1860 UV

CM1860 UV

CM1860 UV

CM1860 UV

Test symbol

-

c_CSA_us

-

-

-

Nominal voltage

100 VAC

120-127 VAC

220-230 VAC

220-230 VAC

240 VAC

±10 %

±10 %

±10 %

±10 %

±10 %

Nominal frequency

50/60 Hz

60 Hz

60 Hz

50 Hz

50 Hz

Power input

1440 VA

1440 VA

1440 VA

1440 VA

1440 VA

Max. start-up current for 5 sec 45 A eff.

45 A eff.

45 A eff.

45 A eff.

45 A eff.

Protective class

I

I

I

I

I

Pollution degree

2

2

2

2

2

Overvoltage inst. category II

II

II

II

II

Heat emission (max.)

1600 J/s

1600 J/s

1600 J/s

1600 J/s

1600 J/s

Refrigeration system

50 Hz/60Hz

Cryochamber

Temperature range

0 °C to -35 °C (+ 3 K/- 3 K)

Cooling time down to -35 °C

max. 6 hrs., at 22 °C ambient temperature

Defrost

Automatic hot gas defrost

1 automatic defrost cycle/24 hours, time-controlled (duration 12 min.)

Refrigeration capacity

690 W

Safety factor

3

Refrigerant

300g ±5g refrigerant R 404A *

Compressor oil

0.6 l EMKARATE RL-22S, ICI *

Quick-freeze shelf

Maximum temperature

- 40 °C (+ 3 K/- 5 K)

Number of freezing stations

8

Defrost

Manual hot gas defrost, time-controlled (duration 12 min.)

Peltier element

Max. temperature difference

17K, at - 35 °C chamber temperature

Number of freezing stations

2

Defrost

In conjunction with the

quick-freeze shelf

according to IEC-1010, UL 3101

according to CECOMAF Liquid temperature 45 °C, evaporation temperature -25 °C

Advertising